Alignment and Resolution Studies of a MARS CT Scanner
Journal article   Peer reviewed

Alignment and Resolution Studies of a MARS CT Scanner

A. Gongadze, A. Zhemchugov, G. Chelkov, D. Kozhevnikov, I. Potrap, M. Demichev, P. Smolyanskiy, R. Abramishvili, S. Kotov, A. P. Butler, …
Physics of particles and nuclei letters, Vol.12(5), pp.725-735
01/09/2015
Handle:
https://hdl.handle.net/10523/38483

Abstract

Physical Sciences Physics Physics, Particles & Fields Science & Technology
This paper describes a procedure of alignment of MARS CT scanner with cone beam geometry. The position of the xray source relative to the detector, the tilt of the detector are measured during alignment. The mechanical precision of gantry movement is determined. Significant improvement in the quality of image reconstruction has been achieved as a result of the calibration. Spatial resolution (PTF) of the scanner is confirmed to be about 30 mu m.